Design and Development of Medical Electronic Instrumentation

Chapter 4 - Electromagnetic Compatibility And Medical Devices: Susceptibility to Magnetic Fields

Susceptibility to Magnetic Fields

IEC-61000-4-8 deals with interference that may be caused on a device by low-frequency
magnetic fields, such as those generated by the power lines. These magnetic fields can produce
jitter on CRT displays, distortion in amplified signals equipment, or false readings in
equipment magnetic or electromagnetic field sensors. As shown in Figure 4.32, the device
under test is placed in the approximate center of a referenced ground plane at a height of
10 cm and is powered and operated in a normal configuration. The magnetic field is
increased to 10 A/m (approximately 125 mG) and is applied to three axes of the device
under test. The field is maintained for a period of approximately 5 minutes for each of
three axes, while the device under test is monitored for any indication of erratic operation.

As shown in Figure 4.33, a test system for design-time evaluation is easy to build. Use
four pieces of 92-cm-long 3/4-in. PVC pipe, four 3/4-in. PVC pipe elbows, and one 3/4-in.
PVC pipe tee to construct the 1 m × 1 m frame for the current loop. Thread No. 18 insulated
copper through the loop to form two complete turns. Solder a flexible twisted-pair cable to
the loop wires where they exit the pipe. R1 will be used to monitor the RMS current flowing
through the coil. The coil is powered by a transformer rated at 24 V at 10 A, which is in
turn powered from the power line through a variac.


Figure 4.31 A simple way of simulating power line voltage dips and interruptions is to use a variac. Although timing an exact 10 ms for a dip is not an easy manual task, good approximations for these dips and interruptions can be achieved when the variac’s output is monitored with an oscilloscope. A small filament transformer provides appropriate isolation for connection with a grounded oscilloscope.


Figure 4.32 Setup for assessing the susceptibility of a medical device to magnetic fields. The magnetic field applied to three axes of the device under test is 10 A/m. This test simulates interference that may be caused on a device by low-frequency magnetic fields, such as those generated by the power lines.


Figure 4.33 A magnetic-field induction generator can be built from a coil of wire inside some PVC pipe. This assembly is driven by a variac–transformer combination to produce the desired magnetic field at the center of the coil.


To use the test system, increase the variac’s output until you read 5.55VRMS across R1.
At this point, 4.5 A should be circulating through the two loops of L1. Correcting for the
difference in magnetic field produced by a square loop versus that of a circular loop, and
applying the Biot–Savart law, the magnetic field at the center of the loop should be
(4.5 A × 2)/0.9m = 10 A/m.

UNLIMITED FREE
ACCESS
TO THE WORLD'S BEST IDEAS

SUBMIT
Already a GlobalSpec user? Log in.

This is embarrasing...

An error occurred while processing the form. Please try again in a few minutes.

Customize Your GlobalSpec Experience

Category: Amplifier and Comparator Chips
Finish!
Privacy Policy

This is embarrasing...

An error occurred while processing the form. Please try again in a few minutes.