Applications of Reference Materials in Analytical Chemistry

4.5: The Preparation of a Calibration Solution Using a CRM

4.5 The Preparation of a Calibration Solution Using a CRM

Some preparation of the CRM is often required before it can be introduced to the instrument and frequently the preparation of a solution of the calibration standard is necessary. Such operations will, in principle, add to the uncertainty of the documented property value, as described in Section 4.4.2 above. The following example shows how the preparation of a solution can affect the uncertainty of the property value.

The CRM LGC 1110 ( p,p ?-DDE) has a certified purity value of 99.6% and an expanded uncertainty (coverage factor, k = 2) of 0.4%. A solution of the material, in hexane, with a nominal concentration of O.1 mg mL ?1 was prepared. Table 4.3 shows the values for the purity of the CRM, the mass of the CRM taken and the volume of hexane solvent used and their estimated standard uncertainties.

Table 4.3: Preparation of a calibration solution using a CRM (1)

Source of uncertainty

Units

Value

u

ru

ru 2

Certified purity

0.996

0.002

0.002008

4.03E 06

Mass of CRM taken

g

0.0114

0.0002

0.017544

0.000308

Volume of hexane

mL

100

0.06

0.0006

3.6E 07

Concentration

mg mL ?1

0.1135

0.0020

0.017669

0.000312

u = standard uncertainty; ru = relative standard uncertainty (= u/value); ru 2 = relative standard uncertainty squared

The concentration of the prepared solution is calculated to be 0.1135 mg mL ?1. The standard uncertainty of this value,...

UNLIMITED FREE
ACCESS
TO THE WORLD'S BEST IDEAS

SUBMIT
Already a GlobalSpec user? Log in.

This is embarrasing...

An error occurred while processing the form. Please try again in a few minutes.

Customize Your GlobalSpec Experience

Category: Laboratory and Calibration Gases
Finish!
Privacy Policy

This is embarrasing...

An error occurred while processing the form. Please try again in a few minutes.