Laminar Flow Gas Flow Meters

Last Updated: April 1, 2025

Description

Laminar Flow Gas Flow Meters are devices designed to measure the volumetric flow rate of gases. They are particularly effective in providing precise flow measurements due to their unique design, which ensures a linear relationship between the differential pressure and the flow rate.

Working Principle

Laminar Flow Gas Flow Meters operate based on the principle of differential pressure. Unlike other differential pressure flowmeters, the pressure drop across a laminar flow element is directly proportional to the volumetric flow rate, rather than the square of the flow rate. This linear relationship allows for more straightforward and accurate flow measurements. The meters require individual calibration due to their complex internal geometry, but once calibrated, they can predict behavior across various operating conditions with high precision. This makes them useful for applications requiring accurate and reliable flow measurements.

Applications

Laminar Flow Gas Flow Meters are used as transfer standards for calibrating other gas flow metering devices. They are suitable for applications where precise flow measurement is critical, such as in laboratory settings or industrial processes that require stringent control over gas flow rates.

Advantages over other Electrical Products

One of the primary advantages of Laminar Flow Gas Flow Meters over other electrical flow measurement devices is their linear relationship between flow rate and differential pressure. This feature simplifies the measurement process and enhances accuracy. Additionally, when calibrated by high-level facilities, these meters can achieve flow rate measurement uncertainties as low as 0.2 to 0.5 percent of reading, which is superior to many other flow measurement technologies.

Limitations

Laminar Flow Gas Flow Meters require specific inlet and outlet sections to function correctly, and they have a minimum Reynolds number requirement for accurate operation. These constraints can limit their use in certain applications where space or flow conditions do not meet these requirements.

Considerations

When considering Laminar Flow Gas Flow Meters, it is important to account for the initial costs associated with their complex design and the need for individual calibration. Operating expenses may be influenced by the need for precise differential pressure instrumentation. However, their durability and high accuracy can offset these costs over time. Maintenance and replacement costs should also be considered, as the meters require careful handling and calibration to maintain their precision.

4 Results
Flow Measurement Sensor -- IF2-C
from Advanced Test Concepts, Inc. (ATC, Inc.)

Medium and low flow range measurement. Portable transfer standard. Calibration of Leak Test orifices and instruments. Measurement of gas mixtures. Pressure up to 500 PSIG. [See More]

  • Meter Technology: Laminar Flow
  • End Fittings: Compression
  • Mounting: In-line
  • Process Media Type: Gas
Meter Tubes
from Daniel Measurement and Control Inc.

Made by experienced code qualified welders, Daniel meter tubes are accurate and dependable instruments that adhere to the highest quality standards. A meter tube consist of an orifice fitting (Daniel Senior, Junior, Simplex or Orifice Flange Union), a length of pipe upstream of the fitting and a... [See More]

  • Meter Technology: Laminar Flow
  • End Fittings: In-line Flanged
  • Mounting: In-line
  • Process Media Type: Liquid; Gas
Mass and Volumetric Flowmeters -- FVL-1600A / FMA-1600A
from OMEGA Engineering, Inc.

The FMA-1600A and FVL-1600A Series mass and volumetric flowmeters use the principle of differential pressure within a laminar flow field to deterime the mass flow rate. A laminar flow element (LFD) insdie the meter forces the gas into laminar (streamlined) flow. Inside this region, the Poiseuille... [See More]

  • Meter Technology: Laminar Flow
  • Mounting: In-line
  • Pipe Diameter: 0.12
  • End Fittings: In-line Threaded
Flow Measurement Sensor -- IF2-HF
from Advanced Test Concepts, Inc. (ATC, Inc.)

High Flow measurements 25 to 10,000 LPM. Precision Transfer Standard to calibrate other flow meters. Constriction/Blockage testing. Measurement of gas mixtures, hydrogen and natural gas options. Pressure range from barometric to 100 psig. [See More]

  • Meter Technology: Laminar Flow
  • End Fittings: In-line Threaded
  • Mounting: In-line
  • Process Media Type: Gas